IntelliVue patient monitors used with 12-lead ECG – risk of ECG trace distortion Specific models and software versions affected. 

Manufactured by Philips Healthcare: Error messages reporting a damaged or incorrectly functioning cable can trigger automatic corrective filtering of the ECG trace, distorting the ST segment, leading to risk of misinterpretation or misdiagnosis. 
 